20
Data Sheet | Version 05.00 R&S®RTC1000 Oscilloscope Specifications year

© Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

  • Upload
    others

  • View
    11

  • Download
    0

Embed Size (px)

Citation preview

Page 1: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Data

She

et |

Vers

ion

05.0

0

R&S®RTC1000OscilloscopeSpecifications

year

RTC1000_dat-sw_en_ 3607-4287-22_v0500_Cover.indd 1 01.10.2018 13:12:54

Page 2: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

2 Rohde & Schwarz R&S®RTC1000 Oscilloscope

CONTENTS Definitions ....................................................................................................................................................................... 3

Base unit .......................................................................................................................................................................... 4

Vertical system .................................................................................................................................................................................. 4

Horizontal system .............................................................................................................................................................................. 4

Acquisition system ............................................................................................................................................................................. 4

Trigger system ................................................................................................................................................................................... 5

Waveform measurements .................................................................................................................................................................. 6

Digital voltmeter ................................................................................................................................................................................. 6

Frequency counter ............................................................................................................................................................................. 6

Component tester .............................................................................................................................................................................. 6

Mask testing ...................................................................................................................................................................................... 7

Waveform maths ............................................................................................................................................................................... 7

Frequency analysis (FFT) .................................................................................................................................................................. 7

Reference signals .............................................................................................................................................................................. 7

Display characteristics ....................................................................................................................................................................... 7

Protocol and logic .............................................................................................................................................................................. 8

Miscellaneous .................................................................................................................................................................................... 8

Input and outputs ............................................................................................................................................................................... 9

General data .................................................................................................................................................................. 10

Options .......................................................................................................................................................................... 11

R&S®RTC-B1 .................................................................................................................................................................................. 11

R&S®RTC-B6 .................................................................................................................................................................................. 12

R&S®RTC-Bxx bandwidth upgrades ................................................................................................................................................ 12

R&S®RTC-K1 .................................................................................................................................................................................. 13

R&S®RTC-K2 .................................................................................................................................................................................. 13

R&S®RTC-K3 .................................................................................................................................................................................. 14

Ordering information .................................................................................................................................................... 16

Page 3: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 3

Definitions General

Product data applies under the following conditions:

Three hours storage at ambient temperature followed by 20 minutes warm-up operation

Specified environmental conditions met

Recommended calibration interval adhered to

All internal automatic adjustments performed, if applicable

Specifications with limits

Represent warranted product performance by means of a range of values for the specified parameter. These specifications are

marked with limiting symbols such as <, ≤, >, ≥, ±, or descriptions such as maximum, limit of, minimum. Compliance is ensured by

testing or is derived from the design. Test limits are narrowed by guard bands to take into account measurement uncertainties, drift

and aging, if applicable.

Specifications without limits

Represent warranted product performance for the specified parameter. These specifications are not specially marked and represent

values with no or negligible deviations from the given value (e.g. dimensions or resolution of a setting parameter). Compliance is

ensured by design.

Typical data (typ.)

Characterizes product performance by means of representative information for the given parameter. When marked with <, > or as a

range, it represents the performance met by approximately 80 % of the instruments at production time. Otherwise, it represents the

mean value.

Nominal values (nom.)

Characterize product performance by means of a representative value for the given parameter (e.g. nominal impedance). In contrast to

typical data, a statistical evaluation does not take place and the parameter is not tested during production.

Measured values (meas.)

Characterize expected product performance by means of measurement results gained from individual samples.

Uncertainties

Represent limits of measurement uncertainty for a given measurand. Uncertainty is defined with a coverage factor of 2 and has been

calculated in line with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into account

environmental conditions, aging, wear and tear.

Device settings and GUI parameters are indicated as follows: “parameter: value”.

Typical data as well as nominal and measured values are not warranted by Rohde & Schwarz.

Page 4: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

4 Rohde & Schwarz R&S®RTC1000 Oscilloscope

Base unit

Vertical system Input channels R&S®RTC1002 2 channels

Input impedance R&S®RTC1002 1 MΩ ± 2 % with 14 pF ± 2 pF (meas.)

Analog bandwidth (–3 dB) R&S®RTC1002 > 50 MHz

R&S®RTC1002 with -B220 option > 70 MHz

R&S®RTC1002 with -B221 option > 100 MHz

R&S®RTC1002 with -B222 option > 200 MHz (≥ 5 mV/div)

R&S®RTC1002 with -B223 option > 300 MHz (≥ 5 mV/div)

Lower frequency limit (–3 dB) at AC coupling < 2 Hz (meas.)

Analog bandwidth limits

(max. –1.8 dB, min. –3.5 dB)

20 MHz (meas.)

Rise time (10 % to 90 %, calculated) R&S®RTC1002 < 7 ns

R&S®RTC1002 with -B220 option < 5 ns

R&S®RTC1002 with -B221 option < 3.5 ns

R&S®RTC1002 with -B222 option < 1.75 ns

R&S®RTC1002 with -B223 option < 1.15 ns

Vertical resolution 8 bit, up to 16 bit with high-resolution

decimation mode

DC gain accuracy maximum operating temperature change of ±5 °C after self-alignment

all input sensitivities ±3 % of full scale

DC measurement accuracy after adequate suppression of

measurement noise by using high-

resolution sampling mode or waveform

averaging

±(DC gain accuracy × |reading +

sensitivity × position setting| + 0.1 div +

1 mV)

Input coupling DC, AC, GND

Input sensitivity 1 mV/div to 10 V/div

Maximum input voltage max. 200 V (Vp), derates at 20 dB/decade

to 5 V (RMS) above 100 kHz

Position range ±15 div

Channel-to-channel isolation

(each channel at same input sensitivity)

input frequency < analog bandwidth > 35 dB (meas.)

Horizontal system Timebase range selectable between 1 ns/div and 100 s/div

Channel deskew ±120 ns

Trigger offset range minimum memory depth/actual sampling rate

maximum 233/actual sampling rate

Modes normal, roll ≥ 50 ms/div

Timebase accuracy after delivery/calibration, at +23 °C ±50 ppm

during calibration interval ±60 ppm

Acquisition system Maximum realtime sampling rate 2 × 1 Gsample/s or 1 × 2 Gsample/s

Memory depth per channel 2 × 1 Msample or 1 × 2 Msample

Acquisition modes refresh first sample in decimation interval

peak detect largest and smallest sample in decimation

interval (1 ns detection)

high resolution average value of all samples in

decimation interval (up to 16 bit)

envelope envelope of acquired waveforms

average average over a series of acquired

waveforms

filter low-pass, adjustable

smooth

Number of averaged waveforms 2 to 1024

Waveform acquisition rate dot display, single channel,

max. waveform rate

up to 10 000 waveforms/s

Page 5: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 5

Trigger system Trigger level range (min) ±15 div from center of screen

Trigger modes auto, normal, single

Hold-off range time auto or 50 ns to 10 s

Trigger types edge, pulse, video, logic, serial bus

Edge trigger trigger events rising edge, falling edge, both edges

sources channel 1, channel 2, logic channels from

D7 to D0 (with R&S®RTC-B1 option),

external trigger input, line

coupling (analog channels, external

trigger input)

DC, AC, auto level,

low pass (attenuates > 5 kHz (meas.)),

HF (attenuates < 30 kHz (meas.)),

noise reject (enlarges trigger hysteresis)

Pulse trigger trigger events pulse width is smaller, greater, equal,

unequal, inside interval, outside interval

min. pulse width 8 ns

max. pulse width 17.1 s

polarity positive, negative

sources channel 1, channel 2, logic channels from

D7 to D0 (with R&S®RTC-B1 option)

Video trigger trigger events selectable line, all lines, even frame,

odd frame, all frames

supported standards PAL, NTSC, SECAM, PAL-M, SDTV 576i,

HDTV 720p, HDTV 1080i, HDTV 1080p

sources channel 1, channel 2

sync pulse polarity positive, negative

Logic trigger trigger events logic condition between active channels

sources channel 1, channel 2, logic channels from

D7 to D0 (with R&S®RTC-B1 option)

state of channels high, low, don’t care

logic between channels and/or

condition true, false

duration condition smaller, greater, equal, unequal, inside

interval, outside interval, timeout

min. duration time 8 ns

max. duration time 17.1 s

Serial bus trigger supported standards

R&S®RTC-K1 option I2C/SPI (two- and three-wire)

R&S®RTC-K2 option UART/RS-232/RS-422/RS-485

R&S®RTC-K3 option CAN/LIN

Trigger sensitivity

with DC, AC, LF reject

input sensitivity ≥ 5 mV/div < 0.8 div (meas.)

input sensitivity < 5 mV/div < 1.5 div (meas.)

with HF reject

all input sensitivities < 1 div (meas.)

with noise rejection

input sensitivity > 5 mV/div < 1.5 div (meas.)

External trigger input input impedance 1 MΩ ± 1 % with 14 pF ± 2 pF (meas.)

maximum input voltage at 1 MΩ max. 100 V (Vp),

derates at 20 dB/decade to 5 V (RMS)

above 100 kHz

trigger level ±5 V

sensitivity 300 mV (Vpp)

input coupling DC, AC

Trigger output (AUX OUT connector) functionality A pulse is generated for every acquisition

trigger event.

output voltage

at high impedance 0 V to 3.0 V

pulse polarity high active

output delay depends on trigger settings

pulse width > 150 ns (trigger event)

> 0.5 µs (mask violation)

Page 6: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

6 Rohde & Schwarz R&S®RTC1000 Oscilloscope

Waveform measurements Automatic measurements measurements on channels,

math waveforms, reference waveforms

burst width, count positive pulses, count

negative pulses, count falling edges,

count rising edges, mean value, RMS,

peak-to-peak, peak+, peak-, frequency,

period, amplitude, crest factor, top level,

base level, pos overshoot, neg overshoot,

pulse width+, pulse witdh-, duty cycle+,

duty cycle-, rise time (80 %, 90 %),

fall time (80 %, 90 %), delay, phase,

standard deviation

measurements on trigger signal trigger period, trigger frequency

implemented by means of six-digit

hardware counter

number of active measurements 6

Cursor measurements measurements on channels,

math waveforms, reference waveforms

voltage (V1, V2, ΔV), time (t1, t2, Δt,

1/Δt), ratio X, ratio Y, pulse and edge

count (pos./neg.), peak values (Vpp, Vp+,

Vpp-), Vmean, VRMS, standard deviation, duty

cycle (pos./neg.), rise/fall time (80 %,

90 %), crest factor, voltage at the cursor

position

functions x and y tracking, coupling of cursors, set

to screen, set to trace, automatic source

Quick measurements function fast overview of measurements from one

channel,

some measurements displayed with result

lines in diagram

sources channel 1, channel 2

measurements displayed in diagram mean value, max. peak, min. peak, rise

time, fall time

numerically displayed measurements RMS, peak-to-peak voltage, period,

frequency, plus 6 automatic

measurements selectable

Marker up to 8 freely positionable markers for

easy navigation

Digital voltmeter Accuracy related to channel settings of voltmeter

source

Measurements DC, AC + DC (RMS), AC (RMS)

Sources channel 1, channel 2

Number of measurements up to 4

Resolution up to 3 digits

Bandwidth > 1 MHz

Frequency counter Measurements frequency, period

Sources trigger signal source (edge, video): line,

channel 1, channel 2, external trigger in

Number of measurements 2

Resolution 5 digits

Frequency range 0. 03 Hz to bandwidth of oscilloscope

(limited by bandwidth of trigger filter)

Component tester Parameters voltage (X), current (Y)

Selectable frequencies 50 Hz, 200 Hz

Component tester output

(AUX OUT connector)

max. output voltage (open circuit) 10 V (Vp) ± 5 %

max. output current (short circuit) 10 mA ± 10 %

reference potential ground

Page 7: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 7

Mask testing Sources channel 1, channel 2

Mask definition acquired waveform with user-defined

tolerance, can be stored and restored

Result statistics completed acquisitions, passed and failed

acquisitions (absolute and in percent),

test duration

Actions on mask violation sound, acquisition stop, screenshot, save

waveform, pulse out (AUX OUT

connector)

Waveform maths Quick math number of math waveforms 1

functions addition, subtraction, multiplication,

division

sources channel 1, channel 2

Mathematics number of formula sets 5

number of equations per set 5

simultaneous display of math waveforms 4

functions addition, subtraction, multiplication,

division, min./max., square, square root,

absolute value, pos./neg. wave,

reciprocal, inverse, log10/ln, derivation,

integration, filter (lowpass/highpass)

sources channel 1, channel 2, math, user defined

constants

Frequency analysis (FFT) Setup parameters center frequency, frequency span,

vertical scale, vertical position

Length 2 ksample to 128 ksample

Window Hanning, Hamming, Blackman,

rectangular, flat top

Waveform arithmetic none, envelope, average (selectable 2 to

512)

Scale dBm, dBV, Veff

Cursor 2 horizontal cursors, previous/next peak

search

Sources channel 1, channel 2

Reference signals Simultaneous display of reference

waveforms

4

Sources analog and digital channels, math,

reference

Display characteristics Diagram types Yt, XY, zoom, FFT, component tester

XY mode

parallel display of XY diagram and

Yt diagrams of input signals for X, Y

Zoom horizontal zoom with fast navigation, split

screen with overview signal and zoomed

signal

FFT mode split screen with overview signal and

dedicated frequency display

Interpolation sin(x)/x, linear, sample & hold

Waveform display lines, dots only

Persistence 50 ms to 9.6 s, infinite

Special display mode inverse brightness, false colors

Diagram grid lines, reticle, none

Virtual screen 20 divisions

Page 8: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

8 Rohde & Schwarz R&S®RTC1000 Oscilloscope

Protocol and logic Bus decode number of bus signals 2 1

bus types parallel, parallel clocked

R&S®RTC-K1 option SSPI, SPI, I2C

R&S®RTC-K2 option UART/RS-232/RS-422/RS-485

R&S®RTC-K3 option CAN, LIN

display types decoded bus, logical signal,

frame table (depends on decoded bus)

data format of decoded bus hex, decimal, binary

Miscellaneous Save/recall device settings save and recall on internal file system or

USB flash drive or on a PC via web

interface

reference waveforms save and recall on internal file system or

USB flash drive or on a PC via web

interface

waveforms save on USB flash drive or download and

save on a PC via web interface

available file formats: BIN, CSV, TXT float

(MSB/LSB first)

screenshots save on USB flash drive or download and

save on a PC via web interface,

available file formats: BMP, PNG, GIF

Print button configurable button, actions on press:

save device settings

save waveforms

save screenshot

save screenshot and setup

Menu languages available menu languages:

English

German

French

Russian

Simplified Chinese

Traditional Chinese

Spanish

Help online help, available languages:

English

German

French

Simplified Chinese

Spanish

1 If a bidirectional bus is used (e.g. UART RX/TX or SPI MOSI/MISO), two bus decoders are occupied.

Page 9: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 9

Input and outputs Front

Channel inputs BNC,

for details see Vertical system

External trigger input trigger in BNC, for details see Trigger system

additional digital channel for level see Trigger system

AUX OUT trigger out for details see Trigger system

mask violation pulse

waveform generator (with R&S®RTC-B6

option only)

BNC, for details see Waveform generator

Probe compensation output signal shape rectangle Vlow = 0 V, Vhigh = 2.4 V (meas.)

frequency 1 kHz and 1 MHz with probe adjust wizard

Pattern source (with R&S®RTC-B6 option

only)

P3 to P0 (with R&S®RTC-B6 option only) 4 lugs, for details see 4-bit pattern

generator

Digital channel inputs D7 to D0 with R&S®RTC-B1 option only

Ground lug connected to ground

USB host interface 1 port, type A plug, version 2.0,

USB drives only

Rear

USB device interface 1 port, type B plug, version 2.0

Ethernet interface 1 port, 1 Gbit

Security slot for standard Kensington style lock

Page 10: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

10 Rohde & Schwarz R&S®RTC1000 Oscilloscope

General data Display

Type 6.5" VGA color display

Resolution 640 × 480 pixel (VGA)

Temperature

Temperature loading operating temperature range +5 °C to +40 °C

storage temperature range –20 °C to +70 °C

Climatic loading +25° C/+40 °C at 85 % rel. humidity

cyclic,

in line with IEC 60068-2-30

Altitude

Operating up to 3000 m above sea level

Nonoperating up to 4600 m above sea level

Mechanical resistance

Vibration sinusoidal 5 Hz to 150 Hz, max. 1.8 g at 55 Hz;

0.5 g from 55 Hz to 150 Hz,

in line with EN 60068-2-6,

MIL-PRF-28800F, 4.5.5.3.2 sinusoidal

vibration, class 3 and 4

random 10 Hz to 300 Hz,

acceleration 1.2 g (RMS),

in line with EN 60068-2-64,

MIL-PRF-28800F, 4.5.5.3.1 random

vibration, class 3 and 4

Shock 40 g shock spectrum,

in line with MIL-STD-810E, method

no. 516.4, procedure I,

MIL-PRF-28800F, 4.5.5.4.1 functional

shock, 30 g, 11 ms, halfsine

Maximum of sound pressure level 30.4 dB (A) at 0.3 m distance

(at +23.6 °C, 931 mbar (hPa), 39 % rel.

humidity), in line with EN ISO 3744

EMC

RF emission in line with CISPR 11/EN 55011 group 1

class A (for a shielded test setup);

the instrument complies with the emission

requirements stipulated by EN 55011,

EN 61326-1 and EN 61326-2-1 class A,

making the instrument suitable for use in

industrial environments

Immunity in line with IEC/EN 61326-1 table 2,

immunity test requirements for industrial

environments 2

Certifications VDE, CCSAUS

Calibration interval 1 year

Power supply

AC supply 100 V to 240 V at 50 Hz to 60 Hz,

100 V to 120 V at 400 Hz

Power consumption max. 25 W

Safety in line with IEC 61010-1, EN 61010-1,

CAN/CSA-C22.2 No. 61010-1,

UL 61010-1

Mechanical data

Dimensions W × H × D 285 × 175 × 140 mm

(11.22 in × 6.89 in × 5.51 in)

Weight without options (nom.) 1.7 kg (3.75 lb)

2 Test criterion is displayed noise level within ±1 div for input sensitivity of 5 mV/div.

Page 11: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 11

Options

R&S®RTC-B1 Mixed signal option, additional 8 logic channels

Vertical system

Input channels 8 logic channels (D7 to D0)

Arrangement of input channels assignment of the logic probes to the

channels D7 to D0

Input impedance 100 kΩ ± 2 % || ~4 pF (meas.) at probe

tips

Maximum input frequency signal with minimum input voltage swing

and hysteresis setting: normal

300 MHz (meas.)

Maximum input voltage ±40 V (Vp)

Minimum input voltage swing hysteresis small 300 mV (Vpp) (meas.)

hysteresis medium 800 mV (Vpp) (meas.)

hysteresis large 1500 mV (Vpp) (meas.)

Threshold groups D7 to D0

Threshold level range –2 V to 8 V in 10 mV steps

predefined CMOS, TTL, ECL

Threshold accuracy ±(100 mV + 3 % of threshold setting)

(meas.)

Comparator hysteresis small, medium, large

Horizontal system

Channel-to-channel skew max. 1 ns (meas.)

Acquisition system

Sampling rate 1 Gsample/s for every channel

Memory depth 1 Msample for every channel

Trigger system see Trigger system

Waveform measurements

Measurement sources all channels from D7 to D0

Automatic measurements positive pulse width, negative pulse width,

period, frequency, burst width, delay,

phase, positive duty cycle, negative duty

cycle, positive pulse count, negative pulse

count, rising edge count, falling edge

count, value at the cursor position

Additional cursor function display of decoded parallel bus value at

the cursor position

Display characteristics

Channel activity display Independent of the oscilloscope

acquisition, the state (stays low, stays

high or toggles) of the channels from D7

to D0 is displayed.

Page 12: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

12 Rohde & Schwarz R&S®RTC1000 Oscilloscope

R&S®RTC-B6 Waveform generator

Resolution 8 bit

Sample rate 978 ksample/s

Amplitude level

high Z 60 mV to 6 V (Vpp)

accuracy 3 % at 1 kHz

DC offset level

high Z ±3 V

accuracy 3 % or ± 25 mV (meas.)

Sine/rectangle frequency 0.1 Hz to 50 kHz

Pulse frequency 0.1 Hz to 10 kHz

Ramp/triangle frequency 0.1 Hz to 10 kHz

4-bit pattern generator

Functions bus signal source 4-bit counter,

programmable 4-bit pattern

Amplitude approx. 2.5 V (Vpp)

Bus signal source SPI, I2C, UART, CAN, LIN

bit rate

UART 9600 bit/s, 115.2 kbit/s, 1 Mbit/s

SPI 100 kbit/s, 250 kbit/s, 1Mbit/s

I2C 100 kbit/s, 400 kbit/s, 1000 kbit/s,

3400 kbit/s

CAN 50 kbit/s, 100 kbit/s, 1 Mbit/s

LIN 9.6 kbit/s, 10.417 kbit/s, 19 kbit/s

4-bit counter frequency 100 mHz to 50 MHz

Squarewave frequency 1 mHz to 500 kHz

duty cycle 1 % to 99 %

Programmable pattern sample time 20 ns to 42 s, up/down

memory depth 2048 sample

pattern idle time 20 ns to 42 s

R&S®RTC-Bxx bandwidth upgrades Option Model Analog bandwidth upgrade

from 50 MHz to

R&S®RTC-B220 R&S®RTC1002 70 MHz

R&S®RTC-B221 R&S®RTC1002 100 MHz

R&S®RTC-B222 R&S®RTC1002 200 MHz

R&S®RTC-B222 R&S®RTC1002 300 MHz

Page 13: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 13

R&S®RTC-K1 I2C triggering and decoding

Bus configuration sources for SCL and SDA channel 1, channel 2, logic channels from

D7 to D0

bit rate up to 10 Mbps

size of address 7 bit or 10 bit

size of data 8 bit

label list associate frame identifier with symbolic ID

Trigger trigger events start, stop, restart, missing acknowledge,

address (7 bit or 10 bit), data, address

and data

offset for trigger on data 0 data byte to 4095 data byte

data pattern width up to 3 sequential data byte

Decode displayed signals bus signal, logic signal or both

color coding of bus signal address, data, start, stop, ACK, NACK;

error and trigger event are displayed in

different colors

displayed format of address hex

displayed format of data ASCII, binary, decimal or hex

SPI triggering and decoding

Bus configuration sources for CS, CLK, data channel 1, channel 2, logic channels from

D7 to D0, extern input (only CS)

bit rate up to 25 Mbps

chip select (CS) active low, active high or missing

(two-wire SPI)

clock (CLK) slope rise or fall

data symbol size 1 bit to 32 bit

idle time for two-wire SPI < 1 ms

Trigger trigger events start of frame, end of frame, bit number,

data pattern

selectable bit number 0 to 4095

offset for trigger on data pattern 0 to 4095 bit

data pattern size 1 bit to 32 bit

Decode displayed signals bus signal, logic signal or both

color coding of bus signal data, start, stop; error and trigger event

are displayed in different colors

displayed format of data ASCII, binary, decimal or hex

data decoding MSB or LSB first

R&S®RTC-K2 UART/RS-232/RS-422/RS-485 triggering and decoding

Bus configuration source for data channel 1, channel 2, logic channels from

D7 to D0

bit rate 300/600/1200/2400/4800/9600/14400/

19200/28800/38400/56000/57600/

115200 bps

128 kbps/256 kbps/1 Mbps or

user-selectable up to 3 Mbps

end of frame timeout, none

signal polarity idle low, idle high

data symbol size 5 bit to 9 bit

parity none, even or odd

stop bits 1, 1.5 or 2

Trigger trigger events start bit, start of frame, symbol number,

any symbol, pattern of symbols, parity

error, frame error, break

offset for trigger on data symbol 0 to 4095 symbols

data symbol pattern width 1 to floor (32/symbol size) symbols

Decode displayed signals bus signal, logic signal or both

color coding of bus signal data, start, stop; error and trigger event

are displayed in different colors

displayed format of data ASCII, binary, decimal or hex

Page 14: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

14 Rohde & Schwarz R&S®RTC1000 Oscilloscope

R&S®RTC-K3 CAN triggering and decoding

Bus configuration signal type CAN_H, CAN_L

sources channel 1, channel 2, logic channels from

D7 to D0

bit rate 10/20/33.3/50/83.3/100/125/250/500/

1000 kbps or user-selectable in range

from 100 bps to 2 Mbps

sampling point 10 % to 90 % within bit period

label list associate frame identifier with symbolic ID

Trigger trigger events start of frame, frame type, identifier,

identifier + data, error condition (any

combination of CRC error, bit stuffing

error, form error and ACK error)

identifier setup frame type (data, remote or both),

identifier type (11 bit or 29 bit);

condition =, ≠, >, <; identifier selectable

from label list

data setup data pattern up to 8 byte (hex or binary);

condition =, ≠, >, <

Decode displayed signals bus signal, logic signal or both

color coding of bus signal start of frame, identifier, DLC, data

payload, CRC, ACK, end of frame, error

frame, overload frame, CRC error, bit

stuffing error, ACK error

displayed format of data hex, decimal, binary, ASCII

frame table decode results displayed as tabulated list,

errors highlighted in red; three table

positions (top, bottom, full screen); frame

navigation; data export as CSV file

Search search events frame, error, identifier, identifier + data,

identifier + error

frame event setup start of frame, end of frame, overload

frame, error frame, data ID 11 bit, data ID

29 bit, remote ID 11 bit, remote ID 29 bit

error event setup any combination of CRC error, bit stuffing

error, form error and ACK error

identifier setup frame type (data, remote or both),

identifier type (11 bit or 29 bit);

condition =, ≠, >, <; identifier selectable

from label list

data setup data pattern up to 8 byte (hex or binary);

condition =, ≠, >, <

event table search results displayed as tabulated list;

event navigation

Page 15: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 15

LIN triggering and decoding

Bus configuration version 1.3, 2.x or SAE J602; mixed traffic is

supported

bit rate 1.2/2.4/4.8/9.6/10.417/19.2 kbps or user-

selectable in range from 1 kbps to 5 Mbps

polarity active high or active low

label list associate frame identifier with symbolic ID

Trigger source any input channel

trigger events start of frame (sync break), identifier,

identifier + data, wakeup frame, error

condition (any combination of checksum

error, parity error and sync field error)

identifier setup range from 0d to 63d; condition =, ≠, >, <;

identifier selectable from label list

data setup data pattern up to 8 byte (hex or binary);

condition =, ≠, >, <

Decode displayed signals bus signal, logic signal or both

color coding of bus signal frame, frame identifier, parity, data

payload, checksum, error condition

displayed format of data hex, decimal, binary, ASCII

frame table decode results displayed as tabulated list,

errors highlighted in red; three table

positions (top, bottom, full screen); frame

navigation; data export as CSV file

Search search events frame, error, identifier, identifier + data,

identifier + error

frame event setup start of frame, wake up

error event setup any combination of checksum error, parity

error and sync field error

identifier setup range from 0d to 63d; condition =, ≠, >, <;

identifier selectable from label list

data setup data pattern up to 8 byte (hex or binary);

condition =, ≠, >, <

event table search results displayed as tabulated list;

event navigation

Page 16: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

16 Rohde & Schwarz R&S®RTC1000 Oscilloscope

Ordering information Designation Type Order No.

R&S®RTC1000 base model

Oscilloscope, 50 MHz, 2 channels R&S®RTC1002 1335.7500.P02

Base unit (including standard accessories: R&S®RT-ZP03 passive probe per channel, R&S®RTC-B6 waveform generator, power

cord, getting started manual and safety instructions)

Choose your bandwidth upgrade

Upgrade of R&S®RTC1002 to 70 MHz bandwidth R&S®RTC-B220 1335.7300.03

Upgrade of R&S®RTC1002 to 100 MHz bandwidth R&S®RTC-B221 1335.7317.03

Upgrade of R&S®RTC1002 to 200 MHz bandwidth R&S®RTC-B222 1335.7275.03

Upgrade of R&S®RTC1002 to 300 MHz bandwidth R&S®RTC-B223 1335.7323.03

Choose your options

Mixed signal upgrade for non-MSO models, 300 MHz R&S®RTC-B1 1335.7281.03

Waveform generator R&S®RTC-B6 1335.7298.03

I²C/SPI serial triggering and decoding R&S®RTC-K1 1335.7230.03

UART/RS-232/RS-422/RS-485 serial triggering and decoding R&S®RTC-K2 1335.7246.03

CAN/LIN serial triggering and decoding R&S®RTC-K3 1335.7252.03

Application bundle, consists of the following options:

R&S®RTC-K1, R&S®RTC-K2, R&S®RTC-K3, R&S®RTC-B6

R&S®RTC-PK1 1335.7330.03

Choose your additional probes

Single-ended passive probes

300 MHz, 10 MHz, 10:1/1:1, 10 MΩ/1 MΩ, 400 V, 12 pF/82 pF R&S®RT-ZP03 3622.2817.02

500 MHz, 10 MΩ, 10:1, 300 V, 10 pF, 5 mm R&S®RT-ZP05S 1333.2401.02

500 MHz, 10 MΩ, 10:1, 400 V, 9.5 pF R&S®RTM-ZP10 1409.7708.02

38 MHz, 1 MΩ, 1:1, 55 V, 39 pF R&S®RT-ZP1X 1333.1370.02

High voltage single-ended passive probes

250 MHz, 100:1, 100 MΩ, 850 V, 6.5 pF R&S®RT-ZH03 1333.0873.02

400 MHz, 100:1, 50 MΩ, 1000 V, 7.5 pF R&S®RT-ZH10 1409.7720.02

400 MHz, 1000:1, 50 MΩ, 1000 V, 7.5 pF R&S®RT-ZH11 1409.7737.02

Current probes

20 kHz, AC/DC, 10 A/1000 A R&S®RT-ZC02 1333.0850.02

100 kHz, AC/DC, 30 A R&S®RT-ZC03 1333.0844.02

10 MHz, AC/DC, 150 A R&S®RT-ZC10 1409.7750.02

100 MHz, AC/DC, 30 A R&S®RT-ZC20 1409.7766.02

120 MHz, AC/DC, 5 A R&S®RT-ZC30 1409.7772.02

Power supply for current probes R&S®RT-ZA13 1409.7789.02

Active differential probes

100 MHz, 1000:1/100:1, 8 MΩ, 1000 V (RMS), 3.5 pF R&S®RT-ZD01 1422.0703.02

200 MHz, 10:1, 1 MΩ, 20 V diff., 3.5 pF R&S®RT-ZD02 1333.0821.02

Logic probes

Active 8 channel logic probe R&S®RT-ZL03 1333.0715.02

Probe accessories

Feedthrough termination 50 Ω R&S®HZ22 3594.4015.02

Adapter BNC/banana R&S®RT-ZA11 1333.0796.02

Probe pouch R&S®RT-ZA19 1335.7875.02

Choose your accessories

Soft case, for R&S®RTC1002 oscilloscope and accessories R&S®RTC-Z3 1333.0867.02

Rackmount kit R&S®ZZA-RTC1K 1333.0967.02

Page 17: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 17

Warranty

Base unit 3 years

All other items 3 1 year

Options

Extended warranty, one year R&S®WE1 Please contact your local

Rohde & Schwarz sales

office.

Extended warranty, two years R&S®WE2

Extended warranty with calibration coverage, one year R&S®CW1

Extended warranty with calibration coverage, two years R&S®CW2

Extended warranty with accredited calibration coverage, one year R&S®AW1

Extended warranty with accredited calibration coverage, two years R&S®AW2

Extended warranty with a term of one and two years (WE1 and WE2)

Repairs carried out during the contract term are free of charge 4. Necessary calibration and adjustments carried out during repairs are

also covered.

Extended warranty with calibration coverage (CW1 and CW2)

Enhance your extended warranty by adding calibration coverage at a package price. This package ensures that your

Rohde & Schwarz product is regularly calibrated, inspected and maintained during the term of the contract. It includes all repairs 4 and

calibration at the recommended intervals as well as any calibration carried out during repairs or option upgrades.

Extended warranty with accredited calibration (AW1 and AW2)

Enhance your extended warranty by adding accredited calibration coverage at a package price. This package ensures that your

Rohde & Schwarz product is regularly calibrated under accreditation, inspected and maintained during the term of the contract. It

includes all repairs 4 and accredited calibration at the recommended intervals as well as any accredited calibration carried out during

repairs or option upgrades.

3 For options that are installed, the remaining base unit warranty applies if longer than 1 year. Exception: all batteries have a 1 year warranty. 4 Excluding defects caused by incorrect operation or handling and force majeure. Wear-and-tear parts are not included.

Page 18: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

18 Rohde & Schwarz R&S®RTC1000 Oscilloscope

Page 19: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

Version 05.00, October 2018

Rohde & Schwarz R&S®RTC1000 Oscilloscope 19

Page 20: © Rohde & Schwarz; R&S®RTC1000 Oscilloscope · Title © Rohde & Schwarz; R&S®RTC1000 Oscilloscope Author: Rohde & Schwarz Created Date: 10/1/2018 1:12:50 PM

R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG

Trade names are trademarks of the owners

PD 3607.4287.22 | Version 05.00 | October 2018 (sk)

R&S®RTC1000 Oscilloscope

Data without tolerance limits is not binding | Subject to change

© 2017 - 2018 Rohde & Schwarz GmbH & Co. KG | 81671 Munich, Germany

Service that adds value Worldwide Local and personalized Customized and flexible Uncompromising quality Long-term dependability

3607

.428

7.22

05.

00 P

DP

1 e

n

Sustainable product design Environmental compatibility and eco-footprint Energy efficiency and low emissions Longevity and optimized total cost of ownership

Certified Environmental Management

ISO 14001Certified Quality Management

ISO 9001

Regional contact Europe, Africa, Middle East | +49 89 4129 12345 [email protected]

North America | 1 888 TEST RSA (1 888 837 87 72) [email protected]

Latin America | +1 410 910 79 88 [email protected]

Asia Pacific | +65 65 13 04 88 [email protected]

China | +86 800 810 82 28 | +86 400 650 58 96 [email protected]

Rohde & SchwarzThe Rohde & Schwarz electronics group offers innovative solutions in the following business fields: test and mea-surement, broadcast and media, secure communications, cybersecurity, monitoring and network testing. Founded more than 80 years ago, the independent company which is headquartered in Munich, Germany, has an extensive sales and service network with locations in more than 70 countries.

www.rohde-schwarz.com

Rohde & Schwarz trainingwww.training.rohde-schwarz.com

3607428722

RTC1000_dat-sw_en_ 3607-4287-22_v0500_Cover.indd 2 01.10.2018 13:12:54